Harmonic Emission - IEC 1000-3-2 , EN 61000-3-2 , (EN60555-2)
Comply: IEC 61000-3-2 Ed.2.1 :2001 (incl. Amd.14) - IEC 61000-4-7 Ed.1.0 :1991
EMC PARTNER AG, SWITZERLAND
HARCS Setup File :
Operator : Reto Casanova Unit : Gas-Discharge Lamp : 15W Serialnumber : none Bemerkungen : Within the limits for class C
Full Bar : Actual Values
Empty Bar : Maximum Values
Blue : Current , Green : Voltage , Red : Failed
Measurement
Date : 20.08.02 16:13:42 V3.10
Urms = 229.7V Freq = 50.000 Range: 0.5 A
Irms = 0.108A Ipk = 0.301A cf = 2.792
P = 15.46W Pap = 24.79VA pf = 0.624
THDi = 72.7 % THDu = 0.10 % Class C
Test - Time : 3min ( 100 %)
Limit Reference: H1(max)= 0.0757A pf(max)= 0.628
Test completed, Result: PASSED
Order Freq. Iavg Iavg%L Imax Imax%L Limit Status
[Hz] [A] [%] [A] [%] [A]
1 50 0.0756 0.0757
2 100 0.0000 0.0001
3 150 0.0595 91.456 0.0596 91.550 0.0651
4 200 0.0000 0.0001
5 250 0.0377 81.727 0.0377 81.727 0.0462
6 300 0.0000 0.0001
7 350 0.0196 0.0197
8 400 0.0000 0.0001
9 450 0.0150 0.0150
10 500 0.0000 0.0001
11 550 0.0145 0.0145
12 600 0.0000 0.0001
13 650 0.0110 0.0111
14 700 0.0000 0.0002
15 750 0.0088 0.0089
16 800 0.0000 0.0001
17 850 0.0086 0.0087
18 900 0.0000 0.0001
19 950 0.0074 0.0076
20 1000 0.0000 0.0001
21 1050 0.0062 0.0063
22 1100 0.0000 0.0001
23 1150 0.0060 0.0061
24 1200 0.0000 0.0001
25 1250 0.0055 0.0057
26 1300 0.0000 0.0001
27 1350 0.0000 0.0049
28 1400 0.0000 0.0001
29 1450 0.0000 0.0046
30 1500 0.0000 0.0001
31 1550 0.0000 0.0045
32 1600 0.0000 0.0001
33 1650 0.0000 0.0041
34 1700 0.0000 0.0001
35 1750 0.0000 0.0038
36 1800 0.0000 0.0001
37 1850 0.0000 0.0037
38 1900 0.0000 0.0001
39 1950 0.0000 0.0034
40 2000 0.0000 0.0001
Calculation of Individual Harmonic Limits For Class C the Limits are related to H1max which is the Maximum Current of the Fundamental during the test time. H1max = 0.076 A ( measured over test time ) For the H1max calculation the Fundamental Current (H1) is taken over every 16 periods and smoothed with an 1,5second filter. For active power less than 25W the alternative test method is used. Current should begin to flow before 60°, has his last peak before 65°, and should not stop flowing before 90°. 2 no Limit defined 3 0.0651 0.0976 =86% of H1 (0.076A) 4 no Limit defined 5 0.0461 0.0692 =61% of H1 (0.076A) 6 no Limit defined 7 no Limit defined 8 no Limit defined 9 no Limit defined 10 no Limit defined 11 no Limit defined 12 no Limit defined 13 no Limit defined 14 no Limit defined 15 no Limit defined 16 no Limit defined 17 no Limit defined 18 no Limit defined 19 no Limit defined 20 no Limit defined 21 * no Limit defined 22 no Limit defined 23 * no Limit defined 24 no Limit defined 25 * no Limit defined 26 no Limit defined 27 * no Limit defined 28 no Limit defined 29 * no Limit defined 30 no Limit defined 31 * no Limit defined 32 no Limit defined 33 * no Limit defined 34 no Limit defined 35 * no Limit defined 36 no Limit defined 37 * no Limit defined 38 no Limit defined 39 * no Limit defined 40 no Limit defined EUT is PASSED if: - all Average values of the Individual Harmonic Currents (Iavg) are below 100% of the Individual Limits. - all Maximum values of the Individual Harmonic Currents (Imax) are below 150% of the Individual Limits. Exceptions: - Individual Harmonic Currents less than 5mA or less than 0.6% of Irms ( which is 0.006*0.108 = 0.001A) are disregaded.
Definitions of Abbreviations
Urms *** Actual total Voltage in Volt RMS
Irms *** Actual total Current in Ampere RMS
Ipk *** Actual Peak value of the Current in Ampere
cf *** Actual Crest Factor (Ipk/Irms)
P *** Actual Active Power in Watt
Pap *** Actual Apparent Power in VA (Urms*Irms)
pf *** Actual Power Factor (P/Pap)
THDi *** Actual Total Harmonic Current Distortion in %
THDu *** Actual Total Harmonic Voltage Distortion in %
THC *** Actual Total Harmonic Current in Ampere
PHC *** Actual Partial Harmonic Current in Ampere
Individual measurements for 2nd to 40th order:
Iavg Average value of the Individual Harmonic Current
in Ampere RMS
Iavg%L Average value of the Individual Harmonic Current
in percentage of the applicable Limit
Imax Maximum Individual Harmonic Current
in Ampere RMS
Imax%lim Maximum Individual Harmonic Current
in percentage of the applicable Limit
Limit Irms Individual Limit (100%) for the selected Class
in Ampere RMS
General :
- Maximum and Average values are calculatet over the full test-time
- The values marked with "***" are actual values which could vary
during test-time and are taken at the time of protocol printout.
- The individual measurements are taken over every 16 periods and
smoothed with an 1,5second filter.
Notes Harmonics Emission Test on a Gas-Discharge Lamp (15W) ---------------------------------------------------- Procedure: The lamp is warmed up for 15 seconds before test start. The rated power is below 25W, so the alternative test method is used. Result: The lamp is within the limits for Class-C.