General
Test Procedure:
"Each individual capacitor shall be subjected to a maximum of 24 pulses of the same polarity. The time between impulses shall be not less than 10s. If any three successive impulses are shown by the monitor to have had a waveform indicating that no self-healing breakdowns or flashovers have taken place in the capacitor then no further impulses shall be applied and the capacitor shall be counted as passed. If all 24 impulses have been applied to the capacitor and three or more of them are of waveform indicating that no self-healing breakdowns or flashovers have taken occurred then the capacitor shall be counted as passed, but if less than three impulses are of the required waveform then the capacitor shall be counted as a defective."
Waveform monitor:
An oscilloscope
compares the waveshape with a reference waveshape taken from a good sample. If there is
any amplitude difference greater than 10% the capacitor has failed. Further the peak value
of the pulse has to be within +/-10% and the duration within +/-20%.
Test Setup: click here for then Test-Setup
Test Levels
| Test Level | Reference pulse ( good sample ) |
Sample test ( 8 samples under test ) |
Example |
| . | in Port 1 | Port order : 1 to 8 | . |
| 1000V | SETUP | SETUP | HTML |
| 2500V | SETUP | SETUP | HTML |
| 4000V | SETUP | SETUP | HTML |
| 5000V | SETUP | SETUP | HTML |
| 6000V | SETUP | SETUP | HTML |
| 8000V | SETUP | SETUP | HTML |
| . | . | . | . |
Under the EMC-PARTNER
Application Notes you find an example of a test on a
Capacitor: RIFA
PME271M, 600nF, 275Vac ; Test voltage : 2500V
The files are
HTML-files just as generated by TEMA protocol engine.